MRI guidelines for professionals
What you'll find on this page
- MRI guidelines for Cochlear™ Nucleus® System
- MRI guidelines for Cochlear™ Osia® System
- MRI guidelines for Cochlear™ Baha® System
Cochlear™ Nucleus® Implants
Designed for access to MRI scans
Cochlear™ Nucleus® Implants have been designed with MRI access in mind for two decades. Access to MRI scans vary based on the generation of the device with our legacy products having more conditions to follow for scans. Details are outlined in our MRI Guidelines.
Cochlear Nucleus Profile™ Plus, the latest implant offering, provides access to MRI scans at 1.5 T and 3.0 T with magnet in place. The Profile, CI24RE, CI24R, and CI24M implants are approved for MRI scans at 1.5T with the magnet in place with the use of a Cochlear Nucleus MRI Kit, or 3.0T with the magnet surgically removed unless otherwise contraindicated. MRI access is available for the majority of Nucleus implant recipients, but the specific conditions and access differ based on implant type.
Each cochlear implant generation has specific guidelines that need to be followed. The approved scan types for implants are as follows:
Implant Type |
MRI Status at 1.5 T* |
MRI Status at 3.0 T* |
---|---|---|
Profile Plus (CI632, CI624, CI622, CI612) |
Magnet in place |
Magnet in place |
Profile (CI532, CI522, CI512, ABI541) |
Magnet in place with the use of the Nucleus MRI Kit |
Magnet should be surgically removed |
Freedom (CI422, CI24REH, CI24RE(CA), CI24RE(ST), ABI24M) |
Magnet in place with the use of the Nucleus MRI Kit |
Magnet should be surgically removed |
CI24R, CI24M |
Magnet in place with the use of the Nucleus MRI Kit |
Magnet should be surgically removed |
CI11+11+2M |
Magnet in place with the use of the Nucleus MRI Kit |
Contraindicated for MRI |
CI22M with removable magnet |
Magnet in place with the use of the Nucleus MRI Kit |
Contraindicated for MRI |
CI22M without removable magnet |
Contraindicated for MRI |
Contraindicated for MRI |
*For full conditions, please refer to the Cochlear Nucleus Implants MRI Guidelines:
- MRI Components Order Form
- Nucleus MRI Guidelines (US)
- Nucleus MRI Guidelines (Canada)
- Nucleus MRI Kit User Guide
- Nucleus MRI Safety Checklist
For further assistance on scanning patients with cochlear implants, call our MRI Information Line at:
1-866-210-9217.
MRI Guidelines for Cochlear™ Osia® System
The Cochlear™ Osia® System was designed with the same technology that has allowed for access to MRI for over two decades in Cochlear Nucleus Implant recipients. This key design feature allows for removal of the magnet.
The Osia System is MRI Conditional at 1.5T with the implant magnet in place and the use of the Cochlear MRI Kit for Osia, or with the implant magnet surgically removed, and at 3.0T with the implant magnet surgically removed.
Implant Type |
MRI Status at 1.5 T** |
MRI Status at 3.0 T** |
---|---|---|
Osia (OSI200 with BI300) |
Magnet in place with use of Cochlear MRI Kit for Osia |
Magnet must be surgically removed |
**For additional details, please refer to the Cochlear Osia MRI Guidelines:
- Osia MRI Guidelines
- Osia MRI Kit User Guide
- Osia MRI Safety Checklist for Professionals
- MRI Components Order Form
For further assistance on scanning patients with Osia System, call our MRI Information Line at:
1-866-210-9217.
MRI Guidelines for Cochlear™ Baha® System
Baha® Connect System
For the Baha Connect System, as long as the sound processor is removed, you can have an MRI scan without any additional risk. There could be a resulting shadow in the image.
Baha® Attract System
The Baha Attract System includes a magnet implanted under the skin, and is MRI conditional for scans at 1.5T. There may be a significant shadow effect for scans of the head. For access to higher strength scans or to reduce the shadow effect, the magnet can be temporarily removed in a straightforward procedure. Baha sound processors should be removed prior to any MRI scan.
Implant Type |
MRI Status at 1.5 T*** |
MRI Status at 3.0 T*** |
---|---|---|
Baha Connect |
May undergo MRI with implant in place |
May undergo MRI with implant in place |
Baha Attract |
May undergo MRI with BI300 implant and BIM400 (internal magnet) in place |
May undergo MRI with BI300 implant in place and BIM400 (internal magnet) must be surgically removed |
***For specific radiology instructions, please refer to the Cochlear Baha System Radiographer’s Instructions for MRI:
- Baha Connect System Radiographer’s Instructions for MRI
- Baha Attract System Radiographer’s Instructions for MRI
For further assistance on scanning patients with Baha system, call our MRI Information Line at:
1-866-210-9217.
Disclaimer
This material is intended for health professionals. If you are a consumer, please seek advice from your health professional about treatments for hearing loss. Outcomes may vary, and your health professional will advise you about the factors which could affect your outcome. Always read the instructions for use. Not all products are available in all countries. Please contact your local Cochlear representative for product information.